iTEMP TMT82 Operating menu and parameter description
Endress+Hauser 109
Options Continuous
Window
Rising
Falling
On change
Factory setting Continuous
Burst trigger level
Navigation Expert → Communication → Burst configuration → Burst trigger level
Prerequisite This parameter can only be selected if the Burst mode option is enabled.
Description Use this function to enter the value which, together with the trigger mode, determines the
time of burst message 1. This value determines the time of the message.
User entry -1.0e
+20
to +1.0e
+20
Factory setting -1.0e
+20
Burst min period
Navigation Expert → Communication → Burst configuration → Burst min period
Prerequisite This parameter can only be selected if the Burst mode option is enabled.
Description Use this function to enter the minimum time span between two burst commands of burst
message X. The value is entered in the unit 1/32 milliseconds.
User entry 500 to [value entered for the maximum interval in the Burst max period parameter] in
whole numbers
Factory setting 1000
Burst max period
Navigation Expert → Communication → Burst configuration → Burst max period
Prerequisite This parameter can only be selected if the Burst mode option is enabled.
Description Use this function to enter the maximum time span between two burst commands of burst
message X. The value is entered in the unit 1/32 milliseconds.
User entry [Value entered for the minimum interval in the Burst min period parameter] to 3600000
in whole numbers