Mounting Liquiphant FTL62 HART
12 Endress+Hauser
A0042281
 2 Handling the device during transportation
5 Mounting
Mounting instructions
Any orientation for device with short pipe up to approx. 500 mm (19.7 in)
Vertical orientation from above for device with long pipe
Minimum distance between the fork tip and the tank wall or pipe wall: 10 mm (0.39 in)
A0042153
 3 Installation examples in a vessel, pipe or tank
1 Temperature spacer/pressure-tight feedthrough (optional) for tank with insulation and/or high process
temperatures
5.1 Mounting requirements
NOTICE
Scratches or impacts damage the coated surface of the device.
Ensure the device is handled properly and professionally during all mounting work.
5.1.1 Pay attention to the temperature for devices with a PFA
coating (conductive)
The difference in temperature between the outer and inner side of the flange may not
exceed 60 °C (140 °F).
If necessary, use external insulation.