Liquiphant FTL62 HART Maintenance
Endress+Hauser 47
Navigation: System → Device management → Reset device
 For details see the "Description of device parameters" documentation.
11.8.2 Resetting the device via operating keys on the electronic insert
Display
onoff
1
2
A0046338
 26 Operating keys on the electronic insert
1+2 Operating keys for device reset (as-delivered state)
Resetting the device (order configuration)
Press both operating keys simultaneously for at least 12 s.
The device has been reset (as-delivered state).
11.9 Device information
All the device information is contained in the Information submenu.
Navigation: System → Information
 For details see the "Description of device parameters" documentation.
11.10 Firmware history
The firmware version can explicitly be ordered via the product structure. This makes it
possible to ensure the compatibility of the firmware version with an existing or
planned system integration.
11.10.1 Version 01.00.zz
Original software
12 Maintenance
No special maintenance work is required.
12.1 Maintenance tasks
12.1.1 Cleaning
It is not permitted to use the device with abrasive media. Material abrasion on the tuning
fork can result in the device malfunctioning.