Liquiphant FTL62 HART Mounting
Endress+Hauser 15
> 40 (1.57)
A0042205
 8 Installation example for a highly viscous liquid. Unit of measurement mm (in)
5.1.4 Avoid buildup
Use short installation sockets to ensure that the tuning fork projects freely into the
vessel
Leave sufficient distance between the buildup expected on the tank wall and the tuning
fork
A0042206
 9 Installation examples for a highly viscous process medium
5.1.5 Take clearance into consideration
Allow sufficient space outside the tank for mounting, connection and settings involving
the electronic insert.