Liquiphant FTL62 HART Commissioning
Endress+Hauser 31
Device variable Device variable code
Terminal current 5
Terminal voltage 6
The device variables can be queried by a HART
®
master using HART
®
command 9 or
33.
8.2.2 System units
The oscillation frequency is specified in Hz. The temperature can be displayed in °C, °F or K.
9 Commissioning
9.1 Preliminaries
L
WARNING
The settings of the current output are relevant for safety!
Incorrect settings may cause the product to overflow or a pump to run dry.
The setting for the current output depends on the setting in the Assign PV parameter.
After changing the setting of the current output: check the settings for the range
(Lower range value output (LRV) and Upper range value output (URV)) and reconfigure
them if necessary!
9.1.1 As-delivered state
If no customized settings were ordered:
Assign PV parameter Level limit detection (8/16 mA mode)
MAX safety mode
Alarm condition set to min. 3.6 mA
DIP switch for locking in OFF position
Bluetooth switched on
Density range > 0.7 g/cm³
Switching times 0.5 s when the fork is covered and 1.0 s when it is uncovered
HART burst mode switched off
9.2 Function check
Before commissioning the measuring point, check whether the post-mounting and post-
connection checks have been performed:
A
 Checklist in "Post-mounting check" section
A
 Checklist in "Post-connection check" section
9.3 Switching on the device
All configuration tools provide a commissioning assistant that supports the user when
setting the most important configuration parameters (Guidance menu
Commissioning wizard).