Liquiphant FTL62 HART Technical data
Endress+Hauser 51
27.5
(1.08)
40³
(1.57)
A0023713
 29 Plug-in jack M12 IP69. Unit of measurement mm (in)
Plug-in jack M12 IP67
Elbowed 90 °
5 m (16 ft) PVC cable (gray)
Slotted nut Cu Sn/Ni
Body: PUR (black)
Order number: 52010285
27.5
(1.08)
40³
(1.57)
A0022292
 30 Plug-in jack M12 IP67. Unit of measurement mm (in)
15 Technical data
15.1 Input
15.1.1 Measured variable
Level (point level), MAX or MIN safety
15.1.2 Measuring range
Depends on the installation location and the pipe extension ordered
Sensor length:
With plastic coating, maximum 3 m (9.8 ft)
With enamel coating, maximum 1.2 m (3.9 ft)
15.2 Output
15.2.1 Output signal
SIO (FEL60H electronic insert)
8/16 mA (SIO) with superimposed digital communication protocol HART, 2-wire