Liquiphant FTL62 HART Mounting
Endress+Hauser 17
5.2.2 Installation
Aligning the tuning fork using the marking
The correct alignment allows the medium to flow easily off the tuning fork and prevents
the formation of buildup
A0042207
12 Markings to align the tuning fork
Installing in pipes
• Flow velocity up to 5 m/s with a viscosity of 1 mPa⋅s and density of 1 g/cm
3
(SGU).
Check for correct functioning in the event of other process medium conditions.
• The flow will not be significantly impeded if the tuning fork is correctly aligned and the
marking is pointing in the direction of flow.
• The marking is visible when installed.
A0042208
13 Installation in pipes (take fork position and marking into consideration)
Aligning the cable entry
A0042214
14 Housing with external locking screw and drip loop
The locking screw is not tightened when the device is delivered.
1. Loosen the external locking screw (maximum 1.5 turns).