Liquiphant FTL62 HART Diagnostics and troubleshooting
Endress+Hauser 43
11.4 Adapting the diagnostic information
The event level can be configured:
Navigation: Diagnostics → Diagnostic settings → Configuration
11.5 Queued diagnostic messages
The display alternates between the queued diagnostic messages and the measured value.
Queued diagnostic messages can also be displayed in the Active diagnostics parameter.
Navigation: Diagnostics → Active diagnostics
11.6 Diagnostic list
All the diagnostic messages that are currently queued can be displayed in the Diagnostic
list submenu.
Navigation: Diagnostics → Diagnostic list
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
004 Sensor defective 1. Restart device
2. Replace electronics
3. Replace device
F Alarm
007 Sensor defective 1. Check fork
2. Replace device
F Alarm
042 Sensor corroded 1. Check fork
2. Replace device
F Alarm
049 Sensor corroded 1. Check fork
2. Replace device
M Warning
1)
061 Sensor electronics faulty Replace electronics F Alarm
062 Sensor connection faulty 1. Check Main to sensor
connection
2. Replace electronics
F Alarm
081 Sensor initialization faulty 1. Restart device
2. Contact service
F Alarm
Diagnostic of electronic
201 Electronics faulty 1. Restart device
2. Replace electronics
F Alarm
203 HART Device Malfunction Check device specific diagnosis. S Warning
204 HART Electronic Defect Check device specific diagnosis. F Alarm
242 Firmware incompatible 1. Check software
2. Flash or change main
electronic module
F Alarm
252 Module incompatible 1. Check if correct electronic
module is plugged
2. Replace electronic module
F Alarm
263 Electronic incompatible Check electronic module type F Alarm
270 Main electronics defective Replace main electronics F Alarm
272 Main electronics faulty 1. Restart device
2. Contact service
F Alarm