Diagnostics and troubleshooting Liquiphant FTL62 HART
40 Endress+Hauser
Login via SmartBlue not possible
Possible cause: Device is being put into operation for the first time
Remedial action: Enter the user name ("admin") and the password (device serial number)
Possible cause: There is insufficient power available.
Remedial action: Increase the supply voltage.
Device cannot be operated via SmartBlue
Possible cause: Incorrect password entered
Remedial action: Enter the correct password
Possible cause: Forgotten password
Remedial action: Use the operating key on the electronic insert to reset the password or
contact Endress+Hauser Service (www.addresses.endress.com)
Possible cause: Operator user role has no authorization
Remedial action: Change to the Maintenance user role
11.1.2 Error - SmartBlue operation with Bluetooth®
Device is not visible in the live list
Possible cause: No Bluetooth connection available
Remedial action: Enable Bluetooth in the field device via display or software tool and/or
in the smartphone/tablet
Possible cause: Bluetooth signal outside of range
Remedial action: Reduce distance between field device and smartphone/tablet
The connection has a range of up to 25 m (82 ft).
Operating radius with intervisibility 10 m (33 ft)
Possible cause: Geopositioning is not enabled on Android devices or is not permitted for
the SmartBlue app
Remedial action: Activate/allow geopositioning service on Android device for the
SmartBlue app
Device appears in the live list but a connection cannot be established
Possible cause: The device is already connected with another smartphone/tablet via
Bluetooth
Only one point-to-point connection is permitted
Remedial action: Disconnect the smartphone/tablet from the device
Possible cause: Incorrect user name and password
Remedial action: The standard user name is "admin" and the password is the device serial
number indicated on the device nameplate (only if the password was not changed by the
user beforehand)
If you have forgotten the password, use the operating key on the electronic insert to
reset the password or
contact Endress+Hauser Service (www.addresses.endress.com)
Connection via SmartBlue not possible
Possible cause: Incorrect password entered
Remedial action: Enter the correct password, paying attention to lower/upper case
Connection via SmartBlue not possible
Possible cause: Forgotten password
Remedial action: Use the operating key on the electronic insert to reset the password or
contact Endress+Hauser Service (www.addresses.endress.com)
11.1.3 Additional tests
If no clear cause of the error can be identified or the source of the problem can be both the
device and the application, the following additional tests can be performed:
1. Reset the device to the factory setting.
2. Check the digital point level or sensor frequency (display, HART, .. ).
3. Check that the device concerned is functioning correctly. If the digital value does not
correspond to the anticipated point level or sensor frequency, replace the device.

Table of Contents

Other manuals for Endress+Hauser Liquiphant FTL62

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Liquiphant FTL62 and is the answer not in the manual?

Endress+Hauser Liquiphant FTL62 Specifications

General IconGeneral
BrandEndress+Hauser
ModelLiquiphant FTL62
CategorySwitch
LanguageEnglish

Related product manuals