Inputs: Conductivity Liquistation CSF48
132
Sterilizations
--> "Inputs: General/Frequently occurring functions" section ä 108
Diagnostic behavior
--> "Inputs: General/Frequently occurring functions" section ä 108
Polarization detection (only conductive sensors)
As a result of flow through the electrolyte/electrode interface, reactions take place here which
result in additional voltage. These polarization effects limit the measuring range of conductive
sensors. Sensor-specific compensation increases the level of accuracy at the measuring range
limits.
The controller recognizes the Memosens sensor and automatically uses suitable
compensation. You can view the measuring range limits of the sensor under
Diagnostics/Sensor information/Sensor specifications.
Operation > 80°C <
100nS/cm
Only conductive sensors
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
187 "Operating time"
Operation < 5°C Only inductive sensors
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
188 "Operating time"
Path: Menu/Setup/Inputs/Conductivity/Extended setup/Diagnostics settings/Polarization detetected
Function Options Info
Polarization detetected Options
•On
•Off
Factory setting
Off
Diagnostics code and associated message text:
168 "Polarization"
Path: Menu/Setup/Inputs/Conductivity/Extended setup/Diagnostics settings
Function Options Info