Liquistation CSF48 Programming
45
Settings with the Basic program type with multiple bottles
Sampling mode = "Flow paced VTCV"
Start condition Options
•Immediate
•Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable time.
If Start condition Date/time is selected:
Start date 01.01.2000 to 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program. The
format depends on the option configured under
general settings.
Start time 00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is started.
The format depends on the option configured under
general settings.
Stop condition Options
•Program end
• Continuous
Factory setting
Program end
Program end:
The device stops sampling automatically once it has
run through the set program.
Continuous:
The device runs through the set program continuously
in an infinite loop. Do not forget to empty the bottles.
Assignment bin. output Options
• No binary output config. for
state reporting
• Binary output S:x
Factory setting
No binary output config. for
state reporting
Assignment of the binary output to the program cycle.
Inputs Settings for the inputs can be made as
described in the "Inputs" section.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Flowmeter input Options
• No flow input configured
• Binary input S:x
• Current input S:x
Factory setting
No flow input configured
Select the flow input.
The binary input or the current input must be
configured for this function.
Only the inputs configured as a flow input are
displayed.
Sampling interval (for version
with vacuum pump or
peristaltic pump)
1.000 to 9999.000 m³
Factory setting
10.000 m³
Set the sampling interval. The unit and the number of
decimal places are displayed as configured under
Setup/Inputs.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info