Programming Liquistation CSF48
62
4.3.2 Programming example: Program switchover
The following example explains how to program a program change with flow-paced sampling
as the daily average sample from Monday to Friday and time-paced sampling to a composite
container on Saturday and Sunday.
Via the "Bottle change" menu item, the bottle change can be configured after a certain time or number of samples
if bottle distribution with more than one bottle has been selected and either dynamic or static bottle assignment
has been selected.
If Bottle assignment Statical bottle assignment is selected:
Bottle assignment table
Select a bottle and assign it a subprogram.
Path: Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program
Standard
Program name TDVT 6+20
Bottle configuration 6x + 1x
Bottle volume left 3000 ml
Bottle volume right 20000 ml
Start condition Date/time
Start date DD.MM.YYYY e.g. 03.06.2010
Start time HH:MM:SS e.g. 12:00:00
Stop condition Continuous
Setup subprogram
Programpart MO - FR
"SAVE" - saves the "MO - FR" subprogram
Setup subprogram
Programpart SAT - SUN
"SAVE" - saves the "SAT - SUN" subprogram
"ESC" - the main program is displayed
Bottle assignment Statical bottle assignment
Setup subprogram: MO - FR
Path: Menu/Setup/Sampling programs