Liquistation CSF48 Programming
61
New bottle at disable Options
•No
•Yes
Factory setting
Yes
Bottle synchronization Options
•None
1. bottle change time
1. Time of change + bottle
number
External BC sync input
Factory setting
None
Specific bottles can be assigned specific filling times
with the bottle synchronization function.
For example, bottle 1 is to be filled from midnight to
2 a.m., bottle 2 from 2 a.m. to 4 a.m. etc..
None
The time of sampling and bottle change are not syn-
chronized.
1. bottle change time
Sampling starts with the first bottle. The changeover
to the next bottles is synchronized.
1. Time of change + bottle number
A specific fill time is assigned to each bottle.
External BC sync input
The system changes to the next bottle when an exter-
nal signal is received. The external signal first has to
be configured via the binary input. The binary input
can then be selected as the source.
Assignment bin. output Options
No binary output config. for
state reporting
Binary output S:x
Factory setting
No binary output config. for
state reporting
Assignment of the binary output to the program
cycle.
Use "SAVE" to save the subprogram setup. Then press "ESC" to return to the main program.
A prompt to save the program appears if you have not yet saved the subprogram. You can avoid saving the program by
pressing "ESC".
Inputs Settings for the inputs can be made as descri-
bed in the "Inputs" section.
Bottle assignment
(only possible with multiple
bottles)
This menu item
appears when more
than one bottle is
available, regardless of
the number of subpro-
grams.
Options
No bottle assignment
Dynamical bottle assign-
ment
Statical bottle assignment
Factory setting
Dynamical bottle assignment
No bottle assignment:
Each subprogram fills the same bottle until the bottle
is full. All the subprograms then change to the next
bottle.
Only visible if there is more than one subprogram.
Dynamical bottle assignment:
When the subprogram changes, the system switches
to the next empty bottle.
Statical bottle assignment:
A table can be used to assign a subprogram to each
bottle.
Path: Menu/Setup/Sampling programs