EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 75

Endress+Hauser Liquistation CSF48
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquistation CSF48 Programming
75
4.3.4 Programming example: Event program
The following example explains how to program an event program with time-paced sampling
into a composite container and time-paced sampling into individual bottles. The composite con-
tainer is emptied daily.
a0014126s
Abb. 8: Event-controlled sampling
ApH value: 8.0
BpH value
1 Sampling into individual bottles when pH value > 8.0
2 Sampling into composite container when pH value < 8.0
A
B
1
1
2
8.0
Path: Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program
Advanced
Program name TDTE 12+20
Bottle configuration 12x + 1x
Bottle volume left 1000 ml
Bottle volume right 2000 ml
Start condition Immediate
Stop condition Continuous
Bottle assignment Statical bottle assignment
Setup subprogram: New
Programpart TD1
Sampling mode Time paced CTCV
Sampling interval 15 min
Sampling volume 100 ml

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF48

Related product manuals