Programming Liquistation CSF48
74
Bottle synchronization Options
•None
• 1. bottle change time
• 1. Time of change + bottle
number
• External BC sync input
Factory setting
None
Specific bottles can be assigned specific filling times
with the bottle synchronization function.
For example, bottle 1 is to be filled from midnight to
2 a.m., bottle 2 from 2 a.m. to 4 a.m. etc..
None
The time of sampling and bottle change are not syn-
chronized.
1. bottle change time
Sampling starts with the first bottle. The changeover
to the next bottles is synchronized.
1. Time of change + bottle number
A specific fill time is assigned to each bottle.
External BC sync input
The system changes to the next bottle when an exter-
nal signal is received. The external signal first has to
be configured via the binary input. The binary input
can then be selected as the source.
Assignment bin. output Options
• No binary output config. for
state reporting
• Binary output S:x
Factory setting
No binary output config. for
state reporting
Assignment of the binary output to the program
cycle.
Use "SAVE" to save the subprogram setup. Then press "ESC" to return to the main program.
Inputs Settings for the inputs can be made as descri-
bed in the "Inputs" section.
Bottle assignment
(only possible with multiple
bottles)
This menu item does
not appear unless 2 or
more bottles are
available.
Options
• No bottle assignment
• Dynamical bottle
assignment
• Statical bottle assignment
Factory setting
No bottle assignment
No bottle assignment:
Each subprogram fills the same bottle until the bottle
is full. All the subprograms then change to the next
bottle.
Dynamical bottle assignment:
When the subprogram changes, the system switches
to the next bottle.
Statical bottle assignment:
A table can be used to assign a subprogram to each
bottle.
Via the "Bottle change" menu item, the bottle change can be configured after a certain time or number of samples
if more than 1 subprogram is available and either dynamic or static bottle assignment has been selected.
If Bottle assignment Statical bottle assignment is selected:
Bottle assignment table
Select a bottle and assign it a subprogram.
Path: Menu/Setup/Sampling programs
Function Options Info