Liquistation CSF48 Programming
73
Sample at enable
(not for single sample
and sampling table)
Options
•No
•Yes
Factory setting
Yes
Specify whether the first sample should be taken
directly when the subprogram is enabled.
Sample at disable
(not for single sample
and sampling table and
also not for "Immediate"
and event)
Options
•No
•Yes
Factory setting
No
Specify whether a sample should be taken when the
subprogram is disabled.
Deactivation Options
Bottles full
Enable invalid
Deactivation with event
Factory setting
Enable invalid
Select the disable function of the subprogram:
Bottles full
The subprogram is disabled once all the assigned
bottles have been filled.
Enable invalid
Disable via limit value
Deactivation with event
New parameter can be defined
Bottle change mode Options
•No
•Yes
Factory setting
Yes
No:
The bottle is changed following a disable/enable
Yes:
When the cycle is finished, the system continues fil-
ling the last bottle.
Synchronize samplings Options
•To subprogram start
•To clock
Factory setting
To subprogram start
To subprogram start
The intervals defined in the sampling mode are
enabled when the subprogram is started.
To clock
The intervals defined in the sampling mode are
enabled after a specific time.
For example, if 30 min is entered this means that the
interval is only activated at a time of xx:30.
--> You configure this time in the "Offset
synchronization" menu item.
Path: Menu/Setup/Sampling programs