Programming Liquistation CSF48
72
Range lower value Range of adjustment and fac-
tory setting
Depends on the measured value
Operating mode="In range check" or "Out of range
check"
• The event is triggered if the range lower value +
hysteresis is exceeded for the switch-on duration.
• The event is reset again if the range upper value -
hysteresis is undershot for the duration of the
switch-off delay at least.
Range upper value
Hysteresis Range of adjustment and fac-
tory setting
Depends on the measured value
The hysteresis is the difference between the
switch-on point and the switch-off point if values,
which cause the limit switch to pick up, become closer
or move further apart. It is needed to ensure a stable
switching behavior.
Start delay 0 to 9999 s
Factory setting
0 s
Synonyms: pick-up and drop-out delay
Switch off delay
Delta value Range of adjustment and fac-
tory setting
Depends on the measured value
Operating mode="Change rate"
The event is triggered if the measured value changes
by at least the delta value (both positive and
negative) within the set delta time. The event is
deleted as soon as the rate of change is lower than
the set value and the auto confirmation time has
elapsed.
Delta time 00:01 to 23:59
Factory setting
01:00
Auto Confirm 00:01 to 23:59
Factory setting
00:01
If Enable subprogram External start is selected:
Activation input Options
• No program part start input
configured
• Binary input S:x
Factory setting
No program part start input
configured
Select the input for start of the subprogram.
The binary input must be configured for this function.
Only the configured inputs are displayed.
If Enable subprogram PROFIBUS DP or Modbus is selected:
Activation signal Options
•None
• %0V DO 01
• %0V DO 02
• %0V DO 03
• %0V DO 04
• %0V DO 05
• %0V DO 06
• %0V DO 07
• %0V DO 08
Factory setting
None
Enable subprogram via fieldbus.
Path: Menu/Setup/Sampling programs
Function Options Info