Liquistation CSF48 Operation
Endress+Hauser 123
Sampling mode Options
•Time paced CTCV
•Flow paced VTCV
• Time/flow paced CTVV
•External signal
Factory setting
Time paced CTCV
Time paced CTCV:
A constant sampling volume is taken at steady
intervals.
Flow paced VTCV:
A constant sampling volume is taken at variable
intervals.
Time/flow paced CTVV(only for version with
peristaltic pump):
A variable sampling volume is taken at steady
intervals.
External signal
A pulse at the binary input starts a sampling cycle
The settings that depend on the sampling mode are listed in the "Program type: Basic" section.
Enable subprogram Options
•Immediate
• Individual dates
• Repeating date
•Interval
• Deactivation
Factory setting
Immediate
Immediate:
The subprogram is enabled immediately.
Individual dates:
Set the start and stop dates for enabling the
subprogram.
Repeating date:
Set the start condition, activity time and repetition
interval for the subprogram.
Interval:
Set the start condition, activity time and inactivity
time for the subprogram.
Deactivation:
Only visible if there is more than one subprogram
If Enable subprogram Individual dates is selected:
Individual dates
Set the start and stop times for the subprogram. Enter a new date via "INSERT". Delete a date via "DELETE". You can
assign a maximum of 25 start and stop dates.
If Enable subprogram Repeating date is selected:
Start condition Options
•No delay
• Date/Time
•Time
Factory setting
No delay
No delay:
The subprogram is started when the program is
enabled.
Date/Time:
Set the start date and start time for enabling the
subprogram.
Time:
Set the start time for enabling the subprogram.
Activity time 00:01 to 99:59 HH:MM
Factory setting
00:01 HH:MM
Specify how long the subprogram should be active in
hours and minutes.
The time to be selected depends on the setting for the
repetition mode.
Path: Menu/Setup/Sampling programs
Function Options Info