EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 125

Endress+Hauser Liquistation CSF48
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquistation CSF48 Operation
Endress+Hauser 125
Sample at enable Options
•No
•Yes
Factory setting
Yes
Specify whether the first sample should be taken
directly when the subprogram is enabled.
For example, with intervals, a sample is taken at the
start of every activation interval.
Sample at disable Options
•No
•Yes
Factory setting
No
Specify whether a sample should be taken when the
subprogram is disabled.
For example, with intervals, a sample is taken at the
end of every activation interval.
New bottle at disable Options
•No
•Yes
Factory setting
Yes
Bottle synchronization Options
•None
1. bottle change time
1. Time of change + bottle
number
External BC sync input
Factory setting
None
Specific bottles can be assigned specific filling times
with the bottle synchronization function.
For example, bottle 1 is to be filled from midnight to
2 a.m., bottle 2 from 2 a.m. to 4 a.m. etc..
None
The time of sampling and bottle change are not
synchronized.
1. bottle change time
Sampling starts with the first bottle. The changeover
to the next bottles is synchronized.
1. Time of change + bottle number
A specific fill time is assigned to each bottle.
External BC sync input
The system changes to the next bottle when an
external signal is received. The external signal first
has to be configured via the binary input. The binary
input can then be selected as the source.
Assignment bin. output Options
No binary output configured
Binary output S:x
Factory setting
No binary output configured
Assignment of the binary output to the program
cycle.
Use "SAVE" to save the subprogram setup. Then press "ESC" to return to the main program.
A prompt to save the program appears if you have not yet saved the subprogram. You can avoid saving the program by
pressing "ESC".
Inputs Settings for the inputs can be made as described in
the "Inputs" section.
Path: Menu/Setup/Sampling programs
Function Options Info

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF48