Liquistation CSF48 Operation
Endress+Hauser 85
Diagnostics settings
Sensor fouling
1)
Warning 0 to 10
Factory setting
7
Indicates maintenance work must be performed on
the conductivity sensors.
If there is conductive fouling between the
conductivity 1 and conductivity 2 electrode, a
diagnostic message is displayed when this level of
fouling is reached.
Alarm 7 to 10
Factory setting
10
If there is conductive fouling between the
conductivity 1 and conductivity 2 electrode, a
diagnostic message is displayed when this level of
fouling is reached.
Pump tube life
2)
Control Options
•Off
•On
Factory setting
On
Indicates the pump hose has to be exchanged.
Warning 10 to 50 h
Factory setting
30 h
When the tube has been in operation for this length
of time, a diagnostic message is displayed to
indicate that the tube should be replaced in time.
Alarm 30 to 200 h
Factory setting
50 h
Totalizer 00-00:00 to 49710-06:28
Factory setting
00-00:00
Operating time of the current pump hose in days,
hours and minutes
Reset Action The tube life counter is reset to 0:00 h.
Process sealing
4)
Control Options
•Off
•On
Factory setting
On
Monitoring of the process seal
Warning 100 to 80000
Factory setting
50000
Enter the number of samples until a warning is
issued.
Alarm 50000 to 1000000
Factory setting
80000
Enter the number of samples until an alarm is
issued.
Totalizer Read only Settings for the sample totalizer
Path: Menu/Setup/General settings
Function Options Info