EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 86

Endress+Hauser Liquistation CSF48
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Liquistation CSF48
86 Endress+Hauser
Reset Action
Armature lock postion
4)
Assembly is set to the lock position.
Sample
temperature(optional)
Temp. control Options
•Off
•On
Factory setting
On
Switch the temperature control of the sample
compartment on or off here.
Economy operation Options
•Off
•On
Factory setting
Off
On:
Cooling is not enabled until the program takes the
first sample. After this, the cooling regulator runs
until the program is restarted.
Sample temperature 2 to 20 ˚C
Factory setting
4 ˚C
Set the sample compartment temperature.
Cooling control Options
Standard operation
Quick cool down
Factory setting
Standard operation
The temperature regulator is switched off for a
certain time if quick cool-down is selected.
Defrosting
The automatic defrosting system is preset at the factory. The following menu items can only be viewed and changed in
the Expert mode.
Mode Options
•Off
•On
Factory setting
On
Configuration of the automatic defrosting function
Interval Options
•Hourly
Daily
•Weekly
•Monthly
Factory setting
Hourly
Select the defrosting interval
Time 00-01:00 to 00-23:59
Factory setting
00-04:00
Path: Menu/Setup/General settings
Function Options Info

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF48

Related product manuals