Liquistation CSF48 Operation
Endress+Hauser 87
10.2.6 Extended setup
Diagnostics settings
The list of diagnostic messages displayed depends on the path selected. There are
device-specific messages, and messages that depend on what sensor is connected.
Duration 00:01 to 02:00
Factory setting
00:05
Path: ... /Extended setup/Diagnostics settings/Diag. behavior (optional)
Function Options Info
List of diagnostic messages Select the message to be changed. Only then can you
make the settings for this message.
Diag. code Read only
Diagnostic message Options
•On
•Off
Factory setting
Depends on the message
You can deactivate or reactivate a diagnostic message
here.
Deactivating means:
• No error message in the measuring mode
• No error current at the current output
Error current Options
•On
•Off
Factory setting
Depends on the message
Decide whether an error current should be output at the
current output if the diagnostic message display is
activated.
If general device errors occur, the error current is
output at all the current outputs.
In the case of channel-specific errors, the error
current is only output at the particular current
output.
Status signal Options
• Maintenance (M)
• Out of specification (S)
•Function check (C)
• Failure (F)
Factory setting
Depends on the message
The messages are divided into different error categories
in accordance with NAMUR NE 107.
Decide whether you want to change a status signal
assignment for your application.
Diag. output Options
•None
• Alarm relay
• Relay 1 to n (depends on
the device version)
Factory setting
None
You can use this function to select a relay output and/or
binary output to which the diagnostic message should be
assigned.
For sensors with the Memosens protocol:
Before you can assign the message to an output, you
must first configure a relay output to "Diagnostics"
(Menu/Setup/Outputs, assign the "Diagnostics" function
and set the Operating mode to "as assigned").