Endress+Hauser Proline Promass F 500 - Configuring the Current Input

Endress+Hauser Proline Promass F 500
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Proline Promass F 500 Modbus RS485
100 Endress+Hauser
Navigation
"Setup" menu → I/O configuration
I/O configuration
I/O module 1 to n terminal numbers
→  100
I/O module 1 to n information
→  100
I/O module 1 to n type
→  100
Apply I/O configuration
→  100
Conversion code
→  100
Parameter overview with brief description
Parameter Description User interface / Selection / 
User entry
Factory setting
I/O module 1 to n terminal numbers Shows the terminal numbers used by the I/O
module.
Not used
26-27 (I/O 1)
24-25 (I/O 2)
22-23 (I/O 3)
20-21 (I/O 4)
*
I/O module 1 to n information Shows information of the plugged I/O
module.
Not plugged
Invalid
Not configurable
Configurable
Fieldbus
I/O module 1 to n type Shows the I/O module type. Off
Current output
*
Current input
*
Status input
*
Pulse/frequency/switch
output
*
Off
Apply I/O configuration Apply parameterization of the freely
configurable I/O module.
No
Yes
No
Conversion code Enter the code in order to change the I/O
configuration.
Positive integer 0
* Visibility depends on order options or device settings
10.3.6 Configuring the current input
The "Current input" wizard guides the user systematically through all the parameters that
have to be set for configuring the current input.
Navigation
"Setup" menu → Current input
Current input 1 to n
Terminal number
→  101