Commissioning Proline Promass F 500 Modbus RS485
102 Endress+Hauser
Navigation
"Setup" menu → Status input
Status input 1 to n
Assign status input
→  102
Terminal number
→  102
Active level
→  102
Terminal number
→  102
Response time status input
→  102
Terminal number
→  102
Parameter overview with brief description
Parameter Description User interface / Selection / 
User entry
Factory setting
Terminal number Shows the terminal numbers used by the
status input module.
Not used
24-25 (I/O 2)
22-23 (I/O 3)
20-21 (I/O 4)
*
Assign status input Select function for the status input. Off
Reset totalizer 1
Reset totalizer 2
Reset totalizer 3
Reset all totalizers
Flow override
Off
Active level Define input signal level at which the
assigned function is triggered.
High
Low
High
Response time status input Define the minimum amount of time the
input signal level must be present before the
selected function is triggered.
5 to 200 ms 50 ms
* Visibility depends on order options or device settings
10.3.8 Configuring the current output
The Current output wizard guides you systematically through all the parameters that have
to be set for configuring the current output.
Navigation
"Setup" menu → Current output
Current output 1 to n
Terminal number
→  103
Signal mode
→  103