EasyManua.ls Logo

Endress+Hauser Proline Promass F 500 - Connecting the Operating Tool

Endress+Hauser Proline Promass F 500
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation options Proline Promass F 500 Modbus RS485
84 Endress+Hauser
8.5.1 Connecting the operating tool
Via Modbus RS485 protocol
This communication interface is available in device versions with a Modbus-RS485 output.
1 2
3
A0029437
 29 Options for remote operation via Modbus-RS485 protocol (active)
1 Control system (e.g. PLC)
2 Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP" or Modbus DTM
3 Transmitter
Service interface
Via service interface (CDI-RJ45)
Proline 500 – digital transmitter
open
press
1
2
3
A0029163
 30 Connection via service interface (CDI-RJ45)
1 Computer with Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for accessing the integrated
device Web server or with "FieldCare", "DeviceCare" operating tool with COM DTM "CDI Communication TCP/IP"
or Modbus DTM
2 Standard Ethernet connecting cable with RJ45 connector
3 Service interface (CDI-RJ45) of the measuring device with access to the integrated Web server

Table of Contents

Related product manuals