Proline Promass F 500 Modbus RS485 Commissioning
Endress+Hauser 133
Options Description
Compare The device configuration saved in the device memory is compared with the current
device configuration of the integrated HistoROM.
Clear backup data The backup copy of the device configuration is deleted from the memory of the
device.
Integrated HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
10.4.7 Using parameters for device administration
The Administration submenu systematically guides the user through all the parameters
that can be used for device administration purposes.
Navigation
"Setup" menu → Advanced setup → Administration
‣
Administration
‣
Define access code
→ 133
‣
Reset access code
→ 134
Device reset
→ 134
Using the parameter to define the access code
Navigation
"Setup" menu → Advanced setup → Administration → Define access code
‣
Define access code
Define access code
→ 133
Confirm access code
→ 133
Parameter overview with brief description
Parameter Description User entry
Define access code Restrict write-access to parameters to protect the configuration
of the device against unintentional changes.
Max. 16-digit character string comprising
numbers, letters and special characters
Confirm access code Confirm the entered access code. Max. 16-digit character string comprising
numbers, letters and special characters