Diagnostics and troubleshooting Proline Promass F 500 Modbus RS485
172 Endress+Hauser
Info number Info name
I1651 Custody transfer parameter changed
I1712 New flash file received
I1725 Sensor electronic module (ISEM) changed
I1726 Configuration backup failed
12.12 Resetting the measuring device
Using theDevice reset parameter (→  134) it is possible to reset the entire device
configuration or some of the configuration to a defined state.
12.12.1 Function scope of the "Device reset" parameter
Options Description
Cancel No action is executed and the user exits the parameter.
To delivery settings Every parameter for which a customer-specific default setting was ordered is reset
to this customer-specific value. All other parameters are reset to the factory
setting.
Restart device The restart resets every parameter whose data are in the volatile memory (RAM)
to the factory setting (e.g. measured value data). The device configuration remains
unchanged.
Restore S-DAT backup Restore the data that are saved on the S-DAT. The data record is restored from the
electronics memory to the S-DAT.
12.13 Device information
The Device information submenu contains all parameters that display different
information for device identification.
Navigation
"Diagnostics" menu → Device information
Device information
Device tag
→  173
Serial number
→  173
Firmware version
→  173
Device name
→  173
Order code
→  173
Extended order code 1
→  173
Extended order code 2
→  173