Installation Proline Promass F 500 Modbus RS485
32 Endress+Hauser
2. Install the measuring device or turn the transmitter housing so that the cable entries
do not point upwards.
A0029263
6.2.4 Mounting the transmitter housing: Proline 500 – digital
L
CAUTION
Ambient temperature too high!
Danger of electronics overheating and housing deformation.
Do not exceed the permitted maximum ambient temperature .
If operating outdoors: Avoid direct sunlight and exposure to weathering, particularly in
warm climatic regions.
L
CAUTION
Excessive force can damage the housing!
Avoid excessive mechanical stress.
The transmitter can be mounted in the following ways:
Post mounting
Wall mounting
Post mounting
L
WARNING
Excessive tightening torque applied to the fixing screws!
Risk of damaging the plastic transmitter.
Tighten the fixing screws as per the tightening torque: 2 Nm (1.5 lbf ft).
ø 20…70
( 0.79…2.75)ø
~102 (~ 4.0)
4 x
SW 10
3 x
TX 25
A0029051
 11 Engineering unit mm (in)