Technical data Proline Promass F 500 Modbus RS485
196 Endress+Hauser
Shock resistance Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Impact resistance Rough handling shocks according to IEC 60068-2-31
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
For details, refer to the Declaration of Conformity.
16.9 Process
Medium temperature range
A0031121
T
A
Ambient temperature
T
F
Medium temperature
A Maximum permitted medium temperature at T
A max
= 60 °C (140 °F); higher medium temperatures require a
reduction in the ambient temperature T
F
(derating)
B Maximum permitted ambient temperature at the maximum specified medium temperature of the sensor
Sensor Noninsulated Insulated
A B A B
T
A
T
F
T
A
T
F
T
A
T
F
T
A
T
F
Promass F 500 – digital
1)
60 °C
(140 °F)
130 °C
(266 °F)
55 °C
(131 °F)
150 °C
(302 °F)
60 °C
(140 °F)
90 °C
(194 °F)
45 °C
(113 °F)
150 °C
(302 °F)
Promass F 500 – digital
2)
60 °C
(140 °F)
240 °C
(464 °F)
– – 60 °C
(140 °F)
150 °C
(302 °F)
50 °C
(122 °F)
240 °C
(464 °F)
Promass F 500 – digital
3)
60 °C
(140 °F)
210 °C
(410 °F)
50 °C
(122 °F)
350 °C
(662 °F)
60 °C
(140 °F)
210 °C
(410 °F)
50 °C
(122 °F)
350 °C
(662 °F)
1) Standard version (order code "Meas. Tube Mat., Wetted Parts Surface", option HA, LA, SA, SB, SC)
2) Extended temperature version (order code "Meas. Tube Mat., Wetted Parts Surface", option SD, SE, SF, TH)
3) High-temperature version (order code "Meas. Tube Mat., Wetted Parts Surface", option TT, TU)
Seals
No internal seals
Density 0 to 5 000 kg/m
3