Commissioning Proline Promass F 500 Modbus RS485
132 Endress+Hauser
Navigation
"Setup" menu → Advanced setup → Configuration backup
Configuration backup
Operating time
→  132
Last backup
→  132
Configuration management
→  132
Backup state
→  132
Comparison result
→  132
Parameter overview with brief description
Parameter Description User interface / Selection Factory setting
Operating time Indicates how long the device has been in
operation.
Days (d), hours (h), minutes
(m) and seconds (s)
Last backup Shows when the last data backup was saved
to embedded HistoROM.
Days (d), hours (h), minutes
(m) and seconds (s)
Configuration management Select action for managing the device data in
the embedded HistoROM.
Cancel
Execute backup
Restore
Compare
Clear backup data
Cancel
Backup state Shows the current status of data saving or
restoring.
None
Backup in progress
Restoring in progress
Delete in progress
Compare in progress
Restoring failed
Backup failed
None
Comparison result Comparison of current device data with
embedded HistoROM.
Settings identical
Settings not identical
No backup available
Backup settings corrupt
Check not done
Dataset incompatible
Check not done
Function scope of the "Configuration management" parameter
Options Description
Cancel No action is executed and the user exits the parameter.
Execute backup A backup copy of the current device configuration is saved from the integrated
HistoROM to the memory of the device. The backup copy includes the transmitter
data of the device.
Restore The last backup copy of the device configuration is restored from the device
memory to the device's integrated HistoROM. The backup copy includes the
transmitter data of the device.