Proline Promass F 500 Modbus RS485 Installation
Endress+Hauser 29
RUPTURE DISK
i
2
3
1
DN 8 ( ")...150 (6")!⁄# DN 250 (10")
2
3
A0028903
1 Rupture disk label
2 Rupture disk with 1/2" NPT internal thread with 1" width across flat
3 Transport protection
L
WARNING
Limited functional reliability of the rupture disk.
Danger to persons from escaping fluids!
Do not remove the rupture disk.
When using a rupture disk, do not use a heating jacket.
Make sure that the function and operation of the rupture disk is not impeded through
the installation of the device.
Take precautions to prevent damage and danger to persons if the rupture disk is
actuated.
Observe information on the rupture disk sticker.
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology.
Calibration takes place under reference conditions →  190. Therefore, a zero point
adjustment in the field is generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
To achieve maximum measuring accuracy even with low flow rates
Under extreme process or operating conditions (e.g. very high process temperatures or
very high-viscosity fluids).