Description of Device Functions Proline Prosonic Flow 93T Portable
82 Endress+Hauser
FORMAT
(2602)
!
Note!
This function is not available unless a number was selected in the function ASSIGN
(2600).
Use this function to define the maximum number of places after the decimal point
displayed for the reading in the additional line.
Options
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting
X.XXXX
!
Note!
Note that this setting only affects the reading as it appears on the display, it has no
influence on the accuracy of the system's calculations.
The places after the decimal point as computed by the measuring device cannot
always be displayed, depending on this setting and the engineering unit. In such
instances an arrow appears on the display between the measuring value and the
engineering unit (e.g. 1.2 m3/h), indicating that the measuring system is
computing with more decimal places than can be shown on the display.
DISPLAY MODE
(2603)
!
Note!
This function is only available if VOLUME FLOW BARGRAPH IN % or SIGNAL
BARGRAPH IN % was selected in the function ASSIGN (2600).
Use this function to define the format of the bar graph.
Options
STANDARD (Simple bar graph with 25 / 50 / 75% gradations and integrated sign).
A0001258
SYMMETRY (Symmetrical bar graph for positive and negative directions of flow, with
–50 / 0 / +50% gradations and integrated sign).
A0001259
Factory setting
STANDARD
Functional description
USER INTERFACE INFORMATION LINE CONFIGURATION