Menu Overview and Explanations Installation and Operation Manual - COMTEC
®
/
OXITEC
®
DustEx
2 Doc.-ID: COM_OXI_Dust_11022020
4.2.12 ENOTEC REMOTE
Info
ENOTEC REMOTE is disabled by default. Password and range are not displayed. After activation (only
possible from system access level), the 8-digit password that is requested during the remote connection
setup must be assigned.
The ENOTEC REMOTE password is used for:
authentification and pairing with a smartphone / tablet / laptop / pc.
authentification / login after every connection. Without Authentication / login, device data cannot be read or
modified. The device configuration can also not be altered.
Range limits the transmission power of the ENOTEC REMOTE module. Maximum = 100m, medium = 10m, short =
1m. The actual possible range may vary due to structural factors and the reception strength of the
Smartphone/Tablet.
When an ENOTEC REMOTE connection to the
analyzer is active, the connection is shown in the
upper right corner of the display .
Figure 25 - ENOTEC REMOTE connection active
Info
A maximum of 16 users (smartphones /tablets) can connect to the ENOTEC REMOTE module of an
ENOTEC analyzer.
Should additional users attempt a connection, the connection will fail. In this case, manually switch off the
ENOTEC REMOTE and switch it back on again SYS MENU => System configuration => ENOTEC
REMOTE Settings which will reset the module.
All previously paired users will need to delete their saved connection to the analyzer and re-pair their
devices.
4.2.13 Measuring units
Measuring units can be set for temperature in °C and °F and for pressure in mbar and psi.
4.2.14 Language
Set the language for all text shown on the display. One can choose between English, German, Spanish and Polish.
4.2.15 Change system code
The system code protects the system configuration against unauthorized usage. Settings which may influence the
measurements are also protected.
Info
The system code on delivery is 0000. In this state, entry into the system is granted without
entering the system code. For security reasons, change the code and store it in safety. In case of
loss of the system code, a system reset has to be carried out. The reset process may only be
carried out by trained service technicians. As an option, a 6 digit code is available.