2 Safety Instructions
80113-401_V24023
2023-05
Risks due to the environment
WARNING
Unsuitable temperature or level of humidity during operation
If you operate the unit at an unsuitable temperature or level of hu-
midity, it may sustain damage, fail, or not perform properly.
Danger of injury for patient and medical personnel! Danger of mate-
rial damage.
Operate the unit at a suitable temperature and level of humidity.
You will find the tolerances for temperature and humidity in the
Technical Data.
If other ambient conditions must be observed for operation of the
unit, you will also find them in the Technical Data.
WARNING
Unsuitable temperature or humidity in transit or storage
If you transport or store the unit at an unsuitable temperature or level
of humidity, it may sustain damage and fail.
Danger of injury for patient and medical personnel! Danger of mate-
rial damage.
Transport and store the unit at a suitable temperature and level
of humidity. You will find the tolerances for temperature and hu-
midity in the Technical Data.
If other ambient conditions must be observed for transport and
storage of the unit, you will also find them in the Technical Data.
NOTICE
Insufficient acclimatization time, unsuitable temperature during
acclimatization
If the unit was stored or transported below or above a certain tem-
perature, it will take a certain time and temperature to acclimatize.
If you do not observe the rules, the unit can sustain damage and fail.
Acclimatize the unit according to the rules in the Technical Data.
Do not place the CO
2
bottle close to a heat source. Heaters or air
conditioners are examples of heat sources. Do not actively heat
the CO
2
bottle, e.g. using a fan heater. Do not expose the CO
2
bot-
tle to direct sunlight.
NOTICE
Overheating of the unit due to poor ventilation
If ventilation is poor, the unit can overheat, sustain damage, and fail.
Install the unit in such a way that there is an unobstructed circu-
lation of air around the housing. Installation in confined wall re-
cesses is prohibited.
WARNING
Penetration of liquid into the unit
Risk of electric shock to the medical personnel.
The unit may become damaged and fail.
Make sure no liquid can penetrate the unit.