29 / 80
5 • Installation
80113-401_V24023
2023-05
Chapter 5
Installation
Ambient conditions
Ignition of anesthetics, skin cleansers, and disinfectants in poten-
tially explosive atmospheres
If you place the unit in a potentially explosive atmosphere, anesthet-
ics, skin cleansers, and disinfectants can ignite.
Risk of fire and explosion to the patient and medical personnel! Risk
of damage to property.
Do not place the unit in potentially explosive atmospheres.
Interference with the unit from portable and mobile HF telecom-
munications units (e.g. mobile phones, WLAN units)
Electromagnetic waves emitted by portable and mobile HF telecom-
munications units may affect the unit.
Risk of injury to the patient.
The unit may fail or not perform properly.
When using portable and mobile HF telecommunications units, in-
cluding their accessories, there must be a distance of at least
30 cm between them and the unit and its cords.
Unsuitable temperature or level of humidity during operation
If you operate the unit at an unsuitable temperature or level of hu-
midity, it may sustain damage, fail, or not perform properly.
Danger of injury for patient and medical personnel! Danger of mate-
rial damage.
Operate the unit at a suitable temperature and level of humidity.
You will find the tolerances for temperature and humidity in the
Technical Data.
If other ambient conditions must be observed for operation of the
unit, you will also find them in the Technical Data.
Unsuitable temperature or humidity in transit or storage
If you transport or store the unit at an unsuitable temperature or level
of humidity, it may sustain damage and fail.
Danger of injury for patient and medical personnel! Danger of mate-
rial damage.