2 Safety Instructions
18 / 80
80113-401_V24023
2023-05
WARNING
Damaged, leaky instrument
CO
2
can flow in an uncontrolled fashion and at high pressure into the
patient's body.
Risk of a gas embolism. Risk of death for the patient.
Do not kink flexible instruments. Do not apply mechanical force to
the instruments.
Before each use, visually inspect the instrument and the instru-
ment tube for mechanical damage.
Read the notes on use for the instrument. If it is required to check
the function and leak-tightness of the instrument before use, the
check is described in the notes on use. The check is then manda-
tory.
You must not use a damaged, leaky instrument!
Risk of infection
WARNING
The unit is contaminated
Risk of infection to the patient and medical personnel.
Follow the instructions for cleaning and disinfecting the unit.
WARNING
Alternate use of disinfectant solutions based on different active
ingredients
The agents may affect one another. The disinfection effect of the dis-
infectant solution may be weakened. Risk of infection to the patient
and medical personnel.
The housing plastic may become brittle and break. A color reaction
may occur with plastics.
Do not use these substances alternately.
Fire / explosion
WARNING
Ignition of anesthetics, skin cleansers, and disinfectants in poten-
tially explosive atmospheres
If you place the unit in a potentially explosive atmosphere, anesthet-
ics, skin cleansers, and disinfectants can ignite.
Risk of fire and explosion to the patient and medical personnel! Risk
of damage to property.
Do not place the unit in potentially explosive atmospheres.