2 • Safety Instructions
14 / 80
80113-401_V24023
2023-05
Do not place vessels containing liquids on top of the unit.
If fluid has penetrated into the unit, take the unit out of operation
immediately. You can only start using the unit again once it has
been checked by a service technician.
Electric shock
WARNING
Defective grounded power outlet, power supply network without
proper grounding, inferior-quality power cord, incorrect line volt-
age, multiple power outlets, extension cords
Risk of electric shock to the medical personnel.
Connect the unit / cart to a properly installed grounded power
outlet.
Only connect the unit to a power supply network with proper
grounding.
Use the compatible Erbe power cord (see the Accessories, Com-
patible power cords chapter).
Check the power cord for damage. You must not use a damaged
power cord.
The supply voltage must match the voltage specified on the unit's
rating plate.
Do not use multiple power outlets.
Do not use extension cords.
WARNING
Connection of unit / cart and power supply during cleaning and
disinfection
Risk of electric shock to the medical personnel!
Switch off the unit. Unplug the power plug of the unit.
WARNING
The user touches the patient and the connections at the back of
the unit at the same time.
Patient and medical staff may be put at risk by compensating cur-
rents.
Risk of injury to patient and medical staff. The unit can fail.
Don’t touch the patient and the connections at the back of the unit
at the same time.