6 Working with ERBECRYO 2
54 / 80
80113-401_V24023
2023-05
2. Using a 19-mm open-end wrench, loosen the nut on the gas tube slightly. The
gas is released.
3. Re-tighten the nut on the gas tube (no picture).
Disassembling the unit
Disconnecting an instrument
1. Press the release button. Remove the instrument from the instrument connection.
2. Place the cover on the instrument.
Switching off the unit
1. Use the power switch to switch the unit off.
2. Unplug the power plug.
Disconnecting the waste gas tube
1. Pull back the closure of the waste gas tube.
2. Hold the closure tightly.
3. Pull the waste gas tube off the waste gas nozzle.
4. Release the closure.
Storing the power cord and
footswitch
1. The power cord and footswitch should ideally be stored on the ERBECRYO 2 Cart.
2. Avoid damaging the power cord and footswitch.