7 • Cleaning and disinfection
80113-401_V24023
2023-05
Chapter 7
Cleaning and disinfection
Safety Instructions
Connection of unit / cart and power supply during cleaning and
disinfection
Risk of electric shock to the medical personnel!
Switch off the unit. Unplug the power plug of the unit.
Penetration of liquid into the unit
Risk of electric shock to the medical personnel.
The unit may become damaged and fail.
Make sure no liquid can penetrate the unit.
Do not place vessels containing liquids on top of the unit.
If fluid has penetrated into the unit, take the unit out of operation
immediately. You can only start using the unit again once it has
been checked by a service technician.
The unit is contaminated
Risk of infection to the patient and medical personnel.
Follow the instructions for cleaning and disinfecting the unit.
Alternate use of disinfectant solutions based on different active
ingredients
The agents may affect one another. The disinfection effect of the dis-
infectant solution may be weakened. Risk of infection to the patient
and medical personnel.
The housing plastic may become brittle and break. A color reaction
may occur with plastics.
Do not use these substances alternately.