5 • Installation
80113-401_V24023
2023-05
Installing the unit on an Erbe cart
Please read the User Manual for the cart concerned. There you will find instructions on
how to secure the unit to the cart.
Connecting tubes and the footswitch
Using unsuitable accessories
Risk of injury to patient and medical personnel.
Only use Erbe instruments, footswitches, waste gas tube, gas
tube, gas bottle adapter and switch valve.
Dangerous CO
2
concentration in room
A high concentration of CO
2
in the air can cause suffocation. CO
2
is
heavier than air. If discharge is not controlled, dangerous accumula-
tions can develop close to the ground.
Observe the national regulations regarding concentrations of CO
2
permitted in the workplace.
Connect the Erbe waste gas tube to the waste gas nozzle on the
unit. Direct the waste gas into the anesthetic suction system or
the open air. This procedure is recommended particularly for ac-
tivation over longer periods.
If it is not possible to use the Erbe waste gas tube, ensure that the
room is adequately ventilated: Enable air to circulate by opening
windows and doors or by switching on the air conditioner.
Check the seal of all connections once the bottle valve has been
opened. No gas may escape. If there are leaks, check that all con-
nections are secure. If you cannot rectify the leak, close the bottle
valve and inform Technical Service.