5 Installation
80113-401_V24023
2023-05
Electrical connection
WARNING
Defective grounded power outlet, power supply network without
proper grounding, inferior-quality power cord, incorrect line volt-
age, multiple power outlets, extension cords
Risk of electric shock to the medical personnel.
Connect the unit / cart to a properly installed grounded power
outlet.
Only connect the unit to a power supply network with proper
grounding.
Use the compatible Erbe power cord (see the Accessories, Com-
patible power cords chapter).
Check the power cord for damage. You must not use a damaged
power cord.
The supply voltage must match the voltage specified on the unit's
rating plate.
Do not use multiple power outlets.
Do not use extension cords.
WARNING
Damaged unit or damaged accessories, modified unit or accesso-
ries
Danger of electric shock to the medical personnel! Danger of damage
to property.
Danger of injury for patient and medical personnel! Danger of mate-
rial damage.
Check the unit and accessories for damage before each use (e.g.,
footswitch, instruments, instrument tubes, gas tube, waste gas
tube, cart).
Check the gas connection, waste gas nozzle and instrument con-
nection in particular for damage.
Close the bottle valve if unwanted freezing or leakage occurs.
You must not use a damaged unit or damaged accessories. Re-
place defective accessories.
If the unit or accessories are damaged, please contact Technical
Service.
Blown power fuses may only be replaced by a competent techni-
cian.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in-
validate liability on the part of Erbe Elektromedizin GmbH.
Electrical connection
Connect the unit to a properly installed, grounded power outlet. Use the compatible
Erbe power cord (see the Accessories, Compatible power cords section). If the unit is
installed on an Erbe cart with auxiliary network outlet: Connect the unit to an auxiliary
network outlet. Connect the unit to the mains power using the cart power cord. Follow
the instructions in the user manual for the cart.
Install the unit such that the power plug can be pulled out without problems.
1. Connect the power cord to the power connection at the back of the unit.