9 Technical Data
64 / 80
80113-401_V24023
2023-05
Ambient conditions for operation of unit
Temp eratu re +15 °C to +40 °C
Relative humidity 30% - 75%, non-condensing
Maximum operating altitude 4000 m above sea level
Acclimatization of the unit
If the unit has been stored or transported at temperatures below +15 °C or above +40 °C, the unit will require approx. 3 hours
to acclimatize at room temperature.
Gas supply
Gas bottles Recommended: Gas bottle with 10 kg capacity
Room temperature when operating the gas bottle Recommended: between +15 °C and +25 °C
Storing the gas bottle Up to max. +50 °C
Acclimatizing the gas bottle If gas bottles are stored at below +15 °C and above +25 °C,
the acclimatization time at room temperature is approx. 8
hours.
Bottle type Gas bottles without dip tube
Standards
Classification according to EC Directive 93/42/EEC II b
Classification according to Regulation (EU) 2017/745 II b
Protection class as per EN 60 601-1 I
Type as per EN 60 601-1 Defibrillation-proof type CF applied part