DMTA-20102-01EN, Rev. 7, October 2022
Chapter 348
3.4.3 Charging a Battery
The maximum ambient temperature for OmniScan X3 battery charging is 30 °C.
Above this temperature, the battery will not charge, although the OmniScan X3 flaw
detector can be operated up to a maximum ambient temperature of 45 °C.
The OmniScan X3 DC power adaptor (P/N: OMNI-A-AC [U8767093]) is designed to
power the OmniScan X3 and charge the lithium-ion batteries only (P/N: OMNI-A-
BATT [U8760010]). Do not attempt to charge any other type of battery or use any
other chargers/adaptors (except those recommended by Evident—for example, the
optional external battery charger P/N: OMNI-A-CHRG [U8767077]). Doing so might
cause an explosion or injury. Do not attempt to power or charge other electronic
equipment with the DC power adaptor (P/N: OMNI-A-AC [U8767093]) or external
battery charger (P/N: OMNI-A-CHRG [U8767077]), unless specifically instructed in
the manual. Misuse of the DC power adaptor or external battery charger can cause
other batteries and/or instruments to explode, which could lead to serious injury or
death.
The lithium-ion batteries are not fully charged when the OmniScan X3 is shipped. You
must charge the batteries for two to three hours before using the battery power to
operate the OmniScan X3.
To charge a battery
Connect the OmniScan X3 flaw detector to the DC power adaptor that is
connected to an appropriate AC power source.
OR
Insert a battery into the optional external battery charger (P/N: OMNI-A-CHRG
[U8767077]) that is connected to an appropriate AC power source.