Communicating Through TCP/IP over Telnet
FTB-500 427
Internal Commands of the TCP/IP over Telnet Protocol
CONNECT LINS
Syntax: CONNECT LINS<Unit_Number><Module_Slot_Number>
You must specify the unit number and the slot number identifying the
module to which the session will connect.
This command allows to connect to different modules through TCP/IP. You
can connect to multiple modules from a single session.
If the command is not executed successfully, the possible reasons could
be:
The module is already connected to a different client session.
The module is not present at the specified position.
The provided information does not correspond to a valid LINS.
Note: For backward compatibility reasons, to connect to a single module, you do
not have to use the CONNECT LINS command. A valid instrument
command (for example, Lins10:SOURce:DATA:TELecom:CLEar) for a valid
LINS position will work for a first module. However, you will need to use
the CONNECT LINS command if you want to connect to other modules as
well.