A1
Basic user manual
Finapres Nova
80
Status
Description
Action
NIBP Finger
Connecting
The module is busy
connecting to the system
If after one minute this
message has not
disappeared, turn
device OFF wait a few
seconds and turn it ON
again.
If the error persists, have
the device checked by
Finapres Medical Systems
B.V.
Error
Another error than the
ones described here has
occurred
Turn device OFF wait a
few seconds and turn it
ON again.
If the error persists, have
the device checked by
Finapres Medical Systems
B.V.
Frontend Unit not
connected
Frontend Unit is not
connected
(Re)connect the
Frontend Unit.
Internal communication
error
Internal communication
problem
Turn device OFF wait a
few seconds and turn it
ON again.
If the error persists, have
the device checked by
Finapres Medical
Systems B.V..
Ok
The module is connected
Please reconnect
Frontend Unit
Finapres Nova is in an
error state that can only
be resolved by
reconnection Frontend
Unit while power is on
Reconnect the Frontend
Unit.
If the error persists, have
the device checked by
Finapres Medical
Systems B.V..
HCU
Not connected
The HCU is not connected
to the Frontend Unit
Connect the HCU
Ok
The HCU is connected
Unknown
The Frontend Unit is not
connected or an unknown
error occurred
Connect the Frontend
Unit
NotZeroed
HCU is connected but not
zeroed
Zero the HCU
Zeroed
HCU is connected and
zeroed, height can be
read