A1
Basic user manual
Finapres Nova
81
Status
Description
Action
Zeroing Failed
HCU zero-ing failed
Try again
Try another HCU
If the error persists, have
the device checked by
Finapres Medical Systems
B.V.
The following warnings and error messages will appear in the STATUS BAR.
Warnings
Warnings
Description
Action
Artery contracted
No pressure
measurement due to
contracted artery
Retry.
Try another finger.
Warm the hand.
Communication time out
Internal communication
problem
Turn device OFF wait a
few seconds and turn it
ON again.
If the error persists, have
the device checked by
Finapres Medical Systems
B.V..
Connect Frontend
No Frontend Unit
connected in ready mode
(Re)connect the
Frontend Unit
CPU overloaded
CPU cannot process the
amount of work within
the required interval
Turn device OFF wait a
few seconds and turn it
ON again.
If the error persists, have
the device checked by
Finapres Medical Systems
B.V..
Fan error
System fan failed, might
cause high internal
temperature, see warning
below
If no temperature
warning is given the
device can be used as
normal, otherwise turn
off the device.
No pulse detected
No pulse detected
Retry.
Try another finger.
Warm the hand.
Physiocal error
During Physiocal, an error
was detected, probably
due to movement artifact
Retry.
Minimize finger
movement and
movement artifacts.
Physiocal off
Physiocal is turned off
Reactivate Physiocal by
pressing Physiocal
button in the GUI.
Plethysm low
Problem with
plethysmogram
Retry.
Try another finger.