Version date: 04 April 2016
High internal temperature
Make sure the rear end
of the device has
access to fresh air (i.e.
don't use the device
inside a closed cabinet).
Move device to cooler
location.
NIBP Finger HCU not
zeroed
HCU is connected but not
zeroed
NIBP Finger HCU not
connected
Generated air pressure is
not high enough which
may be caused by an air
hose kinked, leaks, or
deteriorated control
valve.
Check if the finger cuff
air hose is inserted in
the Frontend Unit.
Check if you are using
the correct finger site
(C1 or C2).
Check for leaks or
kinks.
If the error persists, have
the device checked by
Finapres Medical
Systems B.V..
A short-circuit, zero offset
or power supply
unbalance developed.
The error may also occur
if you disconnect the
Frontend Unit e.g. during
a measurement.
Turn off the device
immediately.
Retry power-on briefly
without finger cuff.
Reconnect the Frontend
Unit and try again.
If the error persists, have
the device checked by
Finapres Medical Systems
B.V.
Cuff electrical cable not
connected, or has a faulty
wire.
Insert the connector,
and try again.
Check if you are using
the correct finger site
(C1 or C2).
A wire in the cuff cable
may be broken. Try
another finger cuff.
Odd plethysmogram
detected, probably due to