FLUXUS WW A Menu structure
128 UMFLUXUS_WWV1-1EN, 2018-02-28
input of the upper limit of the physical quantity to be monitored
This display will only be indicated if Alarm Output is activated and MAX
is selected as the switching condition.
input of the lower limit of the physical quantity to be monitored
This display will only be indicated if Alarm Output is activated and MIN
is selected as the switching condition.
input of the limit for the totalizer of the physical quantity to be monitored
This display will only be indicated if Alarm Output is activated and
QUANT. is selected as the switching condition.
input of the hysteresis for the lower or upper limit
This display will only be indicated if Alarm Output is activated and MIN
or MAX is selected as the switching condition.
Program branch Special Funct.
main menu: selection of the program branch Special Funct.
SYSTEM settings
selection of Special Funct.\SYSTEM settings
SYSTEM settings\Set Clock
selection of the displays for the input of the date and the time
SYSTEM settings\Dialogs/Menus
selection of the displays for the activation/deactivation or setting of menu
items in the other program branches
activation of the menu item for the input of the pipe circumference in the
program branch Parameter
x
activation of the menu item for the input of the fluid pressure in the pro-
gram branch Parameter
x
selection of the input mode for the measuring point number in the pro-
gram branch Measuring:
(1234): digits, point, hyphen
(↑↓← ): ASCII editor
x
setting for the display for the input of the transducer distance in the pro-
gram branch Measuring:
user: Only the entered transducer distance will be displayed if the rec-
ommended and the entered transducer distances are identical.
auto: Only the recommended transducer distance will be displayed.
recommended setting: user
x
INIT-resis-
tant
High Limit:
-10.00 m3/h
Low Limit:
-10.00 m3/h
Quantity Limit:
1.00 m3
R1 Hysterese:
1.00 m3/h
par mea opt >SF<
Special Funct.
Special Funct.
SYSTEM settings
SYSTEM settings
Set Clock
SYSTEM settings
Dialogs/Menus
Pipe Circumfer.
off >ON<
Fluid pressure
off >ON<
Meas.Point No.:
(1234) >(↑↓← )<
Transd. Distance
auto >USER<