FLUXUS WW A Menu structure
130 UMFLUXUS_WWV1-1EN, 2018-02-28
input of the time after which the transmitter toggles to the TransitTime
mode if there are no valid measured values in the NoiseTrek mode
range: 0...9999 s
0: without toggling to the TransitTime mode
This display will only be indicated if the automatic toggling between the
TransitTime mode and NoiseTrek mode is activated.
x
input of the time after which the transmitter has to toggle to the
TransitTime mode
range: 0...9999 s
0: without toggling to the TransitTime mode
This display will only be indicated if the automatic toggling between the
TransitTime mode and NoiseTrek mode is activated.
x
input of the time after which the transmitter has to toggle to the NoiseTrek
mode if there are no valid measured values in the TransitTime mode
range: 0...9 999 s
This display will only be indicated if the automatic toggling between the
TransitTime mode and NoiseTrek mode is activated.
x
activation of the display for the difference between the measured and the
expected sound speed of a selected reference fluid during the measure-
ment
x
selection whether the flow velocity is displayed and transmitted with or
without profile correction
x
input of an upper limit of the flow velocity
range: 0.1...25.5 m/s
0 m/s: no detection for outliers
All measured values that are greater than the limit will be marked as outli-
ers.
x
selection of the input of a lower limit for the flow velocity:
• absolut: independent of the flow direction
• sign: dependent on the flow direction
x
activation of the input of a lower limit of the flow velocity
• factory: the default limit of 2.5 cm/s will be used
• user: input of a limit
x
input of the cut-off flow for positive measured values
range: 0...12.7 cm/s (0.127 m/s)
default: 2.5 cm/s (0.025 m/s)
This display will only be indicated if Cut-off Flow = sign and Cut-
off Flow = user is selected.
x
input of the cut-off flow for negative measured values
range: -12.7...0 cm/s
default: -2.5 cm/s
This display will only be indicated if Cut-off Flow = sign and Cut-
off Flow = user is selected.
x
INIT-resis-
tant
NT-Failed |After
TransTime | 60s
NT-Ok,but | Each
check TT | 300s
Keep TT | For
checking | 5s
Flow Velocity
normal >UNCORR.<
Cut-off Flow
absolut >SIGN<
Cut-off Flow
factory >USER<