A Menu structure FLUXUS WW
UMFLUXUS_WWV1-1EN, 2018-02-28 131
input of the cut-off flow for the absolute value of the measured values
range: 0…12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if Cut-off Flow = absolut and
Cut-off Flow = user is selected.
x
input of the max. signal amplification
range: 0...255
0: no limit of the signal amplification
This display will only be indicated if the SuperUser mode is activated.
x
input of the fixed upper limit of the sound speed
range: 0...3 000 m/s
0: the default limit of 1 848 m/s will be used
This display will only be indicated if the SuperUser mode is activated.
x
input of the offset
range: 0...900 m/s
0: the default limit of 300 m/s will be used
This display will only be indicated if the SuperUser mode is activated.
x
activation of the totalizers overflow x
activation of the taking-over of the totalizer values after the measurement
restart
x
input of a time interval after which the transmitter has to detect a long
measurement failure if there are no valid measured values
0: the default value of 30 s is used
This display will only be indicated if the SuperUser mode is activated.
x
input of the number of decimal places of the totalizers
Automatic: dynamic adjustment
Fixed to x digit: 0...4 decimal places
This display will only be indicated if the SuperUser mode is activated.
x
input of the time interval after which the flow totalizer display toggles be-
tween positive and negative flow direction
x
activation of the manual reset of the totalizers
This display will only be indicated if the SuperUser mode is activated.
x
activation of the display of the totalizer sum
This display will only be indicated if the SuperUser mode is activated.
x
activation of the display of the last valid measured value
This display will only be indicated if the SuperUser mode is activated.
x
activation of the turbulence mode x
INIT-resis-
tant
Cut-off Flow
2.5 cm/s
A: Gain threshold
Fail if > 90 dB
A: Bad soundspeed
thresh. 2007 m/s
A: Bad soundspeed
offset: +321 m/s
Quant. wrapping
off >ON<
Quantity recall
off >ON<
Do not total. if
no meas.> 0 s
Total digits
Automatic
Toggle totalizer
0s
3xC clear totals
off >ON<
Show ΣQ
off >ON<
Keep display val
off >ON<
Turbulence mode
off >ON<